Product Description
High Accuracy Testing RTD Thermocouple Equipment Process Calibrator
SPMK518 intelligent process calibrator has powerful function. It meets the requirements of calibration, maintenance and repair of thermal instruments. It is the ideal instruments for both filed and calibration tasks.
Main Feature
1. Measure voltage, current, resistance, frequency, pulse, switch (on and off),also can use RTD / thermocouple measuring temperature, which are mutual interference;
2. Simulate 13 kinds of thermocouple and 11 kinds of RTD measure and output, with cold junction compensation;
3. DC24V power supply isolated from measurement and output;
4. Build in HART Communication;
5. External SPMKSDP Pressure Module as option;
6. AC/DC power supply: Battery power is recommended for better performance.
Detailed Photos
Product Parameters
|
SPMK518 |
SPMK517 |
SPMK516 |
SPMK510 |
Intelligent Process Calibrator |
Multi-function Process Calibrator |
Temperature Calibrator |
Loop Calibrator |
Measurement |
Millivolt voltage |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Voltage |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Current |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Frequency |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Pulse |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Resistance |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
On- off |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
11 RTDs |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
13 thermocouples |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Simulating |
Millivolt voltage |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Voltage |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Current |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Frequency |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Pulse |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Resistance |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Simulate 11 RTDs |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Simulate 13thermocouples |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Other Functions |
HART communication |
AVAILABLE |
NOT AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
External pressure module |
Need to order separately |
Need to order separately |
NOT AVAILABLE |
NOT AVAILABLE |
Thermal book |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
Step programming output |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
24V loop power supply |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Simulating transmitter |
AVAILABLE |
AVAILABLE |
AVAILABLE |
NOT AVAILABLE |
RS232 communication |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Storage |
AVAILABLE |
AVAILABLE |
AVAILABLE |
AVAILABLE |
Technical Data
Electrical Specifications-Measurement (Ambient temperature (20±5)ºC,1 year accuracy)
Signal |
Range |
Accuracy |
Resolution |
Current |
(-30-30)mA |
±(0.01%RD+0.005%F.S) |
0.1μA |
Voltage |
(-30-30)V |
±(0.01%RD+0.005%F.S) |
0.1mV |
mV |
(-75-75)mV |
±(0.01%RD+0.005%F.S) |
0.1μV |
R-3W |
(0-400)Ω (0-4000)Ω |
±(0.02%RD+0.005%F.S) |
1mΩ 10mΩ |
R-4W |
(0-400)Ω (0-4000)Ω |
±(0.01%RD+0.005%F.S) |
1mΩ 10mΩ |
Frequency |
(1-50000)Hz |
±(0.005%RD+0.002%F.S) |
1Hz |
Pulse |
0-999999 |
±1pcs |
1 |
Limit Switch |
If the switch has detective voltage, range is (3-24)V |
Thermocouple |
K E J T N R S B C D G L U |
RTD |
Pt1000(385), Pt500(385), Pt100(385), Pt100(3916), Pt100(3926), Pt10(385), Cu10(427), Cu50(428), Cu100(428), Ni120(672), Ni100(618) |
Electrical Specifications-Source (Ambient temperature (20±5)ºC,1 year accuracy)
Signal |
Range |
Accuracy |
Resolution |
Current |
(0-22)mA |
±(0.02%RD+0.005%F.S) |
0.1μA |
Voltage |
(0-12)V |
±(0.02%RD+0.005%F.S) |
0.1mV |
mV |
(-10-100)mV |
±(0.02%RD+0.005%F.S) |
1μV |
Resistance |
(0-400)Ω |
±(0.02%RD+0.005%F.S) |
1mΩ |
(400-4000)Ω |
±(0.03%RD+0.01%F.S) |
10mΩ |
Frequency |
(0-50000)Hz |
±(0.005%RD+0.002%F.S) |
0.1Hz |
Pulse |
0-999999 |
±1pcs |
1 |
DC24V |
24.0±0.5V |
N/A |
N/A |
Simulating Thermocouple |
K E J T N R S B C D G L U |
Simulating RTD |
Pt1000(385), Pt500(385), Pt100(385), Pt100(3916), Pt100(3926), Pt10(385), Cu10(427), Cu50(428), Cu100(428), Ni120(672), Ni100(618) |
Feature
♣. Measure voltage, current, resistance, frequency, pulse, switch (on and off),also can use RTD / thermocouple measuring temperature, which are mutual interference;
♣. Simulate 13 kinds of thermocouple and 11 kinds of RTD measure and output, with cold junction compensation;
♣. DC24V power supply isolated from measurement and output;
♣. Build in HART Communication;
♣. External SPMKSDP Pressure Module as option;
♣. AC/DC power supply: Battery power is recommended for better performance.
Company Profile
Our Advantages
Packaging & Shipping
Recommend Products
FAQ
1. Why choose us?
Beijing Spake Technology Co., Ltd is one of the most professional manufacturers of calibration instruments in China with 25 years 'experience.
So far, "SPMK" brand calibration instruments have been used in over 70 countries worldwide and enjoy good reputation.
2. What can you buy from us?
We have been successfully manufactured Pressure/ Temperature calibration instruments such as high accuracy Digital Pressure Gauge,Pressure Calibrator, Pressure Calibration Comparator Pumps as well as Process Calibrator and Temperature Calibrator Dry Blocks.
3. Do you have your own factory?
Yes.
4. Delivery time and payment?
Samples Leading Time: 3-7 working days
Delivery Time: Within 7-15 days after payment
Payment: T/T, Western Union, Paypal